Features
High retention efficiency
Low differential pressure
Efficient bacteria and particulate retention
Gamma irradiation
Applications
Vent Filters for fermentors, tanks, etc.
Sterile-grade filtration for compressed air, oxygen, nitrogen etc.
Operating Conditions
Max. Temperature | 80℃ |
Max. Differential Pressure | 0.55 MPa / 25°C |
Materials of Construction
Filter Type | Bricap C/D/L | |
Membrane | Hydrophobic Polyvinylidene Fluoride (PVDF) | |
Core/Cage/End Cap | Polypropylene (PP) | |
Capsule Shell | Polypropylene (PP) | |
O-ring | Polypropylene (PP) | |
Capsule Vent O-ring | Silicone |
|
Removal Ratings | 0.2 μm |
Biological Safety
Bio-compatibility | Meet the requirement of USP<87>In Vitro Cytotoxicity Test; Meet the criteria of the USP<88>Biological Reactivity Test for Class VI-121°C plastics |
Endotoxins | < 0.25 EU/ml |
Effective Filtration Area
Bricap SFU/V | Bricap C/D | Bricap L | Bricap T | Cartridge |
SFU13 : 0.72 cm2 | C01: 290 cm2 | L03: 0.25 m2 | T10: 0.75 m2 | 5" : 0.375 m2 |
SFU25 : 3.4 cm2 | C02: 640 cm2 | L05: 0.375 m2 | T20: 1.5 m2 | 10" : 0.75 m2 |
SFU33 : 4.5 cm2 | C03: 1000 cm2 | L10: 0.75 m2 | T30: 2.25 m2 | 20" : 1.5 m2 |
V42 : 13.8 cm2 | D01: 450 cm2 | L20: 1.5 m2 | 30" : 2.25 m2 | |
V50A : 20 cm2 | L30: 2.25 m2 | 40" : 3.0 m2 | ||
V50D : 20 cm2 |
Sterilization Capability
Sterilization Method/Filter Format | BriScale Disc, Bricale C/ L / T | Bricap D | Cartridge |
Gamma-irradiation | 25-45 Kgy | / | / |
Autoclave (126°C) applies only to Gamma-compatible package | 5 cycle, 30min | / | / |
Autoclave (130°C) applies only to Atoclave package | 25 cycle, 30min | 25 cycle, 30min | 25 cycle, 30min |
SIP (135°C) applies only to cartridge | / | / | 25 cycle, 30min |